1.
International Journal of Laboratory Medicine ; (12): 2953-2954, 2014.
Article in Chinese | WPRIM | ID: wpr-460121

ABSTRACT

Objective To analysis of the thyroid hormone test results of Abbott Architect ci16200 automatic biochemical-im-mune analyzer (Abbott system) and Roche Cobas e601 automated immunoassay analyzer (Roche system) .Methods The thyroid hormone(including T4 ,FT4 ,T3 ,FT3 ,TSH ) levels of 93 cases of serum samples were respectively detected by Abbott system (CMIA) and Roche system (ECLIA) ,and the results were analyzed .Results The thyroid hormone test results of the two systems had a good correlation ,and the coefficients of T4 ,FT4 ,T3 ,FT3 ,and TSH were 0 .960 ,0 .962 ,0 .976 ,0 .900 ,and 0 .999(P<0 .01) . However ,there were significant difference of the detection results of the two systems(P<0 .05) .Conclusion The thyroid hormone test results of Abbott system and Roche system are not comparable .